CHAPTER
Problem Solving / Error Codes 14
29
CODE CAUSE / SOLUTION
Meter has detected an error;
accurate results not possible:
Test Sensor Guide is wet. Clean
inside of Meter (page 25).
Be sure inside is dry before reuse.
Blood applied before Meter is
ready. Repeat test, making sure
blood icon displays.
Test Sensor has been damaged
due to intense heat exposure.
If problem cannot be corrected,
call the Customer Service
Department or 1-800-028-251
(toll free).
Meter has malfunctioned.
Call the Customer Service
Department or 1-800-028-251
(toll free).
28
CHAPTER
14 Problem Solving / Error Codes
CODE CAUSE / SOLUTION
Battery life has ended.
Replace batteries (two lithium
batteries — size CR2016). See
directions — page 26.
Temperature:
— too cold (below 0° C)
— too hot (above 50° C).
Use Meter in an area of correct
operating temperature (10° to
40° C).
Test Sensor removed during the test.
Move Slide back to turn the Meter
off and then repeat the test.
Meter is turned on when the End
Latch is open.
Move Slide back (close) to turn
the Meter off and then close the
End Latch.
{
possible
causes
Page view 18